Monte Carlo Simulation of Siemens Primus plus Linac for ۶ and ۱۸ MV Photon Beams
محل انتشار: مجله فیزیک و مهندسی پزشکی، دوره: 7، شماره: 4
سال انتشار: 1396
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 109
فایل این مقاله در 14 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_JBPE-7-4_003
تاریخ نمایه سازی: 30 دی 1402
چکیده مقاله:
Objective: The aim of the present study is to simulate ۶ MV and ۱۸ MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data.Methods: The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of percentage depth dose (PDD) and beam dose profile for ۵ × ۵ cm۲, ۱۰ × ۱۰ cm۲ and ۲۰ × ۲۰ cm۲ field sizes were measured and compared with simulation results. Moreover, gamma function was used to compare the measurement and simulation data.Results: The results show a good agreement, within ۱%, was observed between the data calculated by the simulations and those obtained by measurement for ۶ MV photon beam, while it was within ۲% for ۱۸ MV photon beam, except in the build-up region for both beams. Gamma index values were less than unity in most data points for all the mentioned energies and fields. To calculate the dose in the phantom, cells were selected in different modes, one of the modes due to the lack of dose gradient and overlapping, produced better results than others produce.Conclusion: There was good settlement between measured and MC simulation values in this research. The simulation programs can be used for photon modes of Siemens Primus Plus linac in conditions in which it is not possible to perform experimental measurements.
کلیدواژه ها:
Radiotherapy ، Siemens Primus plus Linac ، MC Simulation ، ۶ and ۱۸ MV Photon Beams ، Gamma Function
نویسندگان
H Dowlatabadi
Physics Department, School of Sciences, Payame Noor University of Mashhad, Mashhad, Iran
A A Mowlavi
Physics Department, School of Sciences, Hakim Sabzevari University, Sabzevar, Iran
M Ghorbani
Biomedical Engineering and Medical Physics Department, Faculty of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran
S Mohammadi
Physics Department, School of Sciences, Payame Noor University of Mashhad, Mashhad, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :