A study of the validity of the efficiency transfer method to calculate the peak efficiency using c-ray detectorsat extremely large distances

سال انتشار: 1393
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 257

فایل این مقاله در 9 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_JTAP-8-2_004

تاریخ نمایه سازی: 27 مرداد 1397

چکیده مقاله:

The full-energy peak efficiency (FEPE) curvesof the (2 x 2 and 3 x 3 ) NaI (Tl) detectors weremeasured at seven different axial positions from theirsurfaces. The calibration process was done using radioactivepoint sources, which produce a wide energy rangefrom 59.53 up to 1,408.01 keV. This work has beenundertaken to explain the effects of source energy andsourcE-to-detector distance on the detector efficiencycalculations. The study provides an empirical formula tocalculate FEPE based on the efficiency transfer method fordifferent detectors using the effective solid angle ratio atvery large distances and for higher energies. A remarkableagreement between the measured and calculated efficienciesfor the detectors at the sourcE-to-detector distances<35 cm and above that slight difference was observed.

کلیدواژه ها:

Scintillation detectors Full-energy peak efficiency (FEPE) Efficiency transfer method Effective solid angle

نویسندگان

Ahmed M El-Khatib

Physics Department, Faculty of Science, Alexandria University, Alexandria ۲۱۵۱۱, Egypt

Mohamed S Badawi

Physics Department, Faculty of Science, Alexandria University, Alexandria ۲۱۵۱۱, Egypt

Mohamed A Elzaher

Department of Basic and Applied Sciences, Faculty of Engineering, Arab Academy for Science, Technology andMaritime Transport, Alexandria, Egypt

Abouzeid A Thabet

Department of Medical Equipment Technology, Pharos University in Alexandria, Alexandria, Egypt