ON-LINE SPECIATION ANALYSIS OF ANTIMONY VIA HYDRIDE GENERATION TECHNIQUE USING ULTRASONIC NEBULIZERCOUPLED TO ICP-OES
سال انتشار: 1395
نوع سند: مقاله کنفرانسی
زبان: انگلیسی
مشاهده: 657
فایل این مقاله در 5 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
TCPCO03_173
تاریخ نمایه سازی: 16 شهریور 1395
چکیده مقاله:
For the first time, on-line hydride generation using ultrasonic nebulizer was applied for determination of Sb(III) and Sb(V) by inductively coupled plasma optical emission spectrometry (ICPOES). Duo to different kinetics of hydride generation of different oxidation state of antimony, speciation analysis would be possible by this technique. In this work, ultrasonic nebulizer was used for on-line and direct hydride generation as well as for sample for ICP-OES. Parameters affecting on speciation analysis were investigated and under optimal conditions, limits of detection were 0.04 and 0.14 ng mL-1 for Sb(III) and Sb(V), respectively
کلیدواژه ها:
نویسندگان
S.R Yousefi
NFCRS, Nuclear Science and Technology Research Institute Tehran, Iran
E Zolfonoun
NFCRS, Nuclear Science and Technology Research Institute Tehran, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :