Monte Carlo optimization of collimator length for a parallel-hole prompt gamma camera in proton beam range verification
محل انتشار: فصلنامه فیزیک و مهندسی پرتو، دوره: 7، شماره: 4
سال انتشار: 1405
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 41
فایل این مقاله در 7 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_RPE-7-4_007
تاریخ نمایه سازی: 21 آبان 1405
چکیده مقاله:
This study aimed to optimize the collimator length for a parallel-hole prompt gamma camera dedicated to real-time proton beam range verification, achieving an optimal balance between detection sensitivity and spatial accuracy for clinical application. Comprehensive Monte Carlo simulations were conducted using the GATE platform with the QGSP_BIC_HP_EMZ physics list. A cylindrical PMMA phantom was irradiated with a ۹۰ MeV monoenergetic proton beam, and a PG camera model consisting of an LSO scintillator coupled with a lead parallel-hole collimator was simulated. Collimator length was systematically varied from ۴ to ۸ cm, and detected PG events were analyzed within a ۴.۲-۶.۳ MeV energy window. System performance was evaluated using sensitivity (total detected counts) and spatial accuracy (Δ), defined as the absolute difference between true and reconstructed depth of maximum PG emission. The simulated energy spectrum exhibited characteristic photopeaks from carbon and oxygen, and a strong spatial correlation was confirmed between the distal edge of the Bragg peak and the PG emission maximum. Increasing collimator length decreased sensitivity but generally improved spatial accuracy. The optimal trade-off was achieved with a ۷ cm collimator, yielding the best spatial accuracy of Δ = ۲ mm. This study identifies ۷ cm as the optimal collimator length for a parallel-hole PG camera, achieving spatial accuracy comparable to more complex slit- or pinhole-based systems. The findings provide a quantitative design framework and underscore the critical importance of balancing geometric resolution with sufficient count statistics for precise and timely clinical range verification.
کلیدواژه ها:
نویسندگان
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :