Absorbed Dose Measurements for Teletheraphy Prediction of Superficial Dose Using Halcyon Linear Accelerator
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 21، شماره: 4
سال انتشار: 1403
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 86
فایل این مقاله در 7 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-21-4_008
تاریخ نمایه سازی: 13 آبان 1403
چکیده مقاله:
Introduction: Measurement of entrance dose and dose at different depth is essential to avoid over dose and under dose of patients. The aim of this study is to verify the variation in the absorbed dose using a water equivalent material.Material and Methods: The plastic phantom was arranged on the couch of the halcyon linear accelerator by Varian with the farmer ionization chamber inserted and connected to the electrometer. The image of the setup was taken using the High-Quality Single ۱۲۸۰x۱۲۸۰x۱۶ higher on the service mode, to check the alignment with the isocenter. The beam quality TPR۲۰,۱۰ (Tissue phantom ratio) was done to check the beam quality of the machine at a field size of ۱۰ cm x ۱۰ cm. The calibration was done using SAD type set-up at a depth of ۵ cm. This process was repeated for ten consecutive weeks and the values recorded.Results: The results of the beam output for the teletherapy machine were satisfactory and accepted in comparison with the commissioned measurement of ۰.۶۲. The beam quality TPR۲۰,۱۰ (Tissue phantom ratio) was reasonable with respect to the beam quality of the machine at a field size of ۱۰ cm x ۱۰ cm.Conclusion: The results of the beam quality and the absorbed dose rate showed a good consistency over the period of ten weeks with the commissioned measurement value.
کلیدواژه ها:
Linear Accelerator Multi ، Leaf Collimator Volumetric Modulated Arc Therapy Drum Phantom
نویسندگان
Raymond Njinga
Department of Physics, University of Medical Sciences Ondo, Ondo State, Nigeria
Adeneye Samuel
Department of Radio Biology, Radio Diagnosis and Radiotherapy, College of Medicine University of Lagos/NSIA-LUTH Cancer Centre: Idiaraba, Lagos, Nigeria
Akinyode Ajimo
Department of Physics, University of Medical Sciences Ondo, Ondo State, Nigeria
Christopher Olowookere
Department of Physics, University of Medical Sciences Ondo, Ondo State, Nigeria
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :