Over-Response Correction of SNC ۳۵۰p ROOS and PTW Markus Chambers in the Surface Dose and Build-Up Region Dosimetry of ۶MV Photon Beam
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 21، شماره: 4
سال انتشار: 1403
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 90
فایل این مقاله در 8 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-21-4_009
تاریخ نمایه سازی: 13 آبان 1403
چکیده مقاله:
Introduction: The design characteristics of ionization chambers are critical for the accurate measurement of Percentage Surface Dose (PSD) and Percentage Depth Dose (PDD) in the build-up region. This study assesses the over-response correction in PSD and PDD for different field sizes, using Markus and ROOS fixed separation parallel-plate ionization chambers. Material and Methods: The Sun Nuclear Corporation SNC ۳۵۰P ROOS and PTW Markus parallel plate ionization chambers were used to measure ۶ MV photon beam PDD along the central axis in the build-up region.The experiment was performed on a Solid Water® HE phantom with a density of ۱.۰۳۲ g/cm۳.The Measurements were taken at a Source Surface distance (SSD) of ۱۰۰ cm with Field sizes of ۱۰x۱۰cm۲,۲۰x۲۰ cm۲, and ۳۰x۳۰cm۲ and at variable depths. Results: For field sizes of ۱۰x۱۰ cm۲, ۲۰x۲۰ cm۲, and ۳۰x۳۰ cm۲ at a source-to-axis distance (SAD) of ۱۰۰ cm, the PSD values obtained with the Markus chamber, before and after applying Rawlinson's equation for over-response correction, were as follows: ۳۵.۱۱%, ۴۲.۹۲%, ۵۰.۷۲% (before correction) and ۲۸.۳۷%, ۳۶.۱۹%, ۴۴.۰۵% (after correction). Similarly, the PSD values measured using the ROOS chamber, before and after applying over-response correction, were as follows: ۴۱.۶۱%, ۵۰.۵۰%, ۵۸.۰۴% (before correction), and ۳۳.۹۶%, ۴۲.۸۷%, ۵۰.۳۸% (after correction). Conclusion: The application of over-response correction using Rawlinson's equation improved the accuracy of measurements for both chambers, emphasizing the significance of employing appropriate correction techniques to ensure precise dose measurements in the build-up region using ionization chambers.
کلیدواژه ها:
Radiation Dosage Percentage Surface Dose Build ، Up Dose Dosimeters Radiotherapy
نویسندگان
Deepali Patil
Department of Medical Physics,State Cancer Institute, Indira Gandhi Institute of Medical Sciences,Patna, Bihar.
Mukesh Zope
Indira Gandhi Institute Of Medical Sciences ,Sheikhpura ,Patna
DINESH SAROJ
BALCO Medical Center, Vedanta Research Foundation, New Raipur, atal Nagar, Chhattisgarh ۴۹۳۶۶۱
Aslam P. A.
Department of Radiation Oncology, Kiran Multi Super Speciality Hospital, Surat-۳۹۵۰۰۴, Gujarat.
Basil George
Department of Radiation Oncology, STARLIT Cancer Center, Ahmadabad-۳۸۲۳۵۰, Gujarat.
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :