Investigation of the energy response of thermoluminescence dosimeter used for measurement of individual dose-equivalent
محل انتشار: فصلنامه فیزیک و مهندسی پرتو، دوره: 5، شماره: 4
سال انتشار: 1403
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 198
فایل این مقاله در 6 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_RPE-5-4_002
تاریخ نمایه سازی: 14 مهر 1403
چکیده مقاله:
One of the challenges of individual dosimetry in Iran is the absence of standard radiation fields with different energies for calibration. For dosimetry in the photon fields, only the standard fields of Co-۶۰ and Cs-۱۳۷ sources, located in the (SSDL) can be utilized. This study aims to determine energy response of a personal thermoluminescence dosimeter (TLD) designed for determining the personal dose-equivalent, Hp(۱۰), in order to investigate whether the calibration curve in the Co-۶۰ gamma field can be utilized for assessing the dose-equivalent in the photon fields with different energies. To do this, first, the TLD dosimeters (an appropriate plastic badge including a TLD-۱۰۰ chip) are irradiated with a few Hp(۱۰) values using the Co-۶۰ source in the Karaj SSDL. The badges are placed on a water phantom (slab), one meter distant from the source. Then, MCNP۴C code is used to calculate the energy response at ۶۶۲ keV and ۱.۲۵ MeV energies. Next, these responses are validated with the experimental data. Finally, the calculation is carried out for several other energies in the range of ۲۰ keV to ۱.۲۵ MeV. Obtained results show that for the energies more than ۴۰۰ keV, The response of the dosimeter is independent of the photon energy, within ۱۰% uncertainty. On the other hand, for the energies smaller than ۴۰۰ keV, a significant dependence on the energy is observed; such that in ۴۰ keV the response is about ۲ times larger than that at ۱.۲۵ MeV. Consequently, a method to reduce this uncertainty is needed.
کلیدواژه ها:
نویسندگان
Mahdi Norouzkhani
Department of Physics, Shahrood University of Technology, Shahrood, Iran
Amir Moslehi
Radiation Applications Research School, Nuclear Science and Technology Research Institute, AEOI, Tehran, Iran
Vahideh Ataeinia
Management of Nuclear Safeguards, Radiation Protection and HSE, Nuclear Science and Technology Research Institute, AEOI, Tehran, Iran
Moslem Sohani
Department of Physics, Shahrood University of Technology, Shahrood, Iran
Seyed Javad Bizeh
Management of Nuclear Safeguards, Radiation Protection and HSE, Nuclear Science and Technology Research Institute, AEOI, Tehran, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :