Pulse Oximetry as an Alternative Means of Assessing Systolic Blood Pressure in Well Newborns: A Comparison with Two Methods of Blood Pressure Measurement
محل انتشار: مجله بین المللی کودکان، دوره: 12، شماره: 2
سال انتشار: 1403
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 226
فایل این مقاله در 7 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_INJPM-12-2_008
تاریخ نمایه سازی: 30 مرداد 1403
چکیده مقاله:
Background: Blood pressure measurement in newborns using conventional auscultatory and palpatory methods is technically difficult. As a result, different blood pressure measuring devices such as oscillometric and doppler devices have been developed. However, in resource challenged environments, these devices are not widely available unlike pulse oximeters which are relatively cheap and portable. Thus, the objective of this study was to investigate the reliability of using the pulse oximeter in measuring systolic blood pressure in normal, term neonates whilst comparing it to oscillometric and doppler/aneroid blood pressure measuring devices. Methods: This was a prospective, cross-sectional study involving ۱۹۲ healthy terms, appropriate for gestational age neonates who were recruited. Their systolic blood pressures were measured simultaneously using oscillometry, doppler/anaeroid sphygmomanometer and pulse oximetry at ۱۲-hour intervals for the first ۴۸ hours of life. Results: One hundred and eighty-five babies were analyzed. Mean systolic blood pressure readings using doppler with anaeroid sphygmomanometer was ۶۰.۳۵(SD ۲.۲۷), by Oscillometric device, it was ۶۹.۸۳ (SD ۱۲.۵۵) and by pulse oximetry ۵۵.۰۸ (SD ۲.۱۱). Systolic blood pressure readings using a pulse oximeter were significantly lower compared to the other methods and the degree of absolute agreement between the rating instruments was poor especially with oscillometric measurements where the level of agreement further declined with time. Conclusion: Pulse oximetry is a better alternative to oscillometry in systolic blood pressure measurement in newborns especially in resource challenged environments but it cannot be used interchangeably or substituted for doppler anaeroid systolic blood pressure measurement
کلیدواژه ها:
نویسندگان
Frances Okpokowuruk
Department of Paediatrics, University of Uyo Teaching hospital, Uyo, Akwa Ibom state, Nigeria
Ijeoma Ezeasouba
Department of Paediatrics, University of Uyo Teaching hospital, Uyo, Akwa Ibom state, Nigeria
Kevin Bassey
Department of Paediatrics, University of Uyo Teaching hospital, Uyo, Akwa Ibom state, Nigeria
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :