Performance of Iterative Reconstruction in Image Space Algorithm in Combination with Automatic Tube Current Modulation Compared to Filtered Back Projection in Brain CT Scan
محل انتشار: مجله فیزیک و مهندسی پزشکی، دوره: 14، شماره: 4
سال انتشار: 1403
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 361
فایل این مقاله در 10 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_JBPE-14-4_006
تاریخ نمایه سازی: 14 مرداد 1403
چکیده مقاله:
Background: High-quality images with minimum radiation dose are considered a challenge in Computed Tomography (CT) scans.Objective: The current study aimed to assess the efficacy of the Iterative Reconstruction in Image Space (IRIS) algorithm combined with Automatic Tube Current Modulation (ATCM) compared to Filtered Back Projection (FBP) in brain CT scans.Material and Methods: In this cross-sectional study, ۲۰۰ patients underwent to brain CT scan, and images were then reconstructed using both FBP and IRIS. The CT Number (CTN), noise, and Signal-to-Noise Ratio (SNR) were computed for different tissues from CT images. The performance of two algorithms under different exposure conditions was evaluated using a water phantom. Two experienced radiologists assessed the image quality. Volume CT Dose Index (CTDIvol) and Dose Length Product (DLP) were recorded for each scan.Results: FBP reconstruction exhibited higher noise and lower SNR compared to IRIS, both with and without ATCM. Noise levels significantly increased for FBP combined with ATCM. Subjective analysis showed higher performance for IRIS without ATCM compared to other approaches. The mean CTDIvol with and without ATCM was ۲۰.۰۴±۳.۳۳ and ۳۶.۳۷±۴.۶۵ mGy, respectively. In the phantom study, the noise with IRIS remained lower than that with FBP even with a ۴۲% dose reduction. Conclusion: IRIS algorithm can preserve the image quality when radiation dose is significantly reduced by ATCM in brain CT scan. Implementation of IRIS combined with ATCM is recommended for brain CT examinations.
کلیدواژه ها:
نویسندگان
Salman Jafari
Department of Radiology Technology, School of Paramedicine, Hamadan University of Medical Sciences, Hamadan, Iran
Sohrab Kolivand
Department of Radiology, School of Medicine, Hamadan University of Medical Sciences, Hamadan, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :