The Influence of Adaptive Statistical Iterative Reconstruction Algorithm on Image Quality and Volume CT Dose Index in Head CT Scan Compared To Filtered Back Projection
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 20، شماره: 6
سال انتشار: 1402
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 82
فایل این مقاله در 5 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-20-6_002
تاریخ نمایه سازی: 8 بهمن 1402
چکیده مقاله:
Introduction: This study aimed to compare the image quality and volume dose index of head CT scan between two adaptive statistical iterative reconstructions (ASIR) and the filtered back projection (FBP) algorithms.Material and Methods: CT number, noise and signal to noise ratio(SNR) for white matter(WM), gray matter(WM), cerebrospinal fluid(CSF( and skull bone were investigated in brain CT scans of ۶۰ patients. All images were reconstructed by FBP and ASIR ۴۰% algorithms. A water phantom was also used to compare the average CT number; noise, signal-to-noise (SNR), and contrast-to-noise ratio (CNR) between algorithms under different acquisition parameters. Volume computed tomography dose index (CTDIvol) and (dose-length product) DLP were obtained from scanner software. Data were analyzed by T-test and Mann-Whitney statistical test with a significance level of less than ۰.۰۵.Results: Image noise of gray matter, CSF and skull bone was significantly lower for ASIR algorithm (P<۰.۰۵). The difference in SNR for white matter and gray matter was not significant between the two algorithms but it was higher for CSF and bone for ASIR. In phantom study, Image noise, CTDIvol and DLP in both axial and spiral scan modes were higher for FBP algorithm (P<۰.۰۵). In addition, there was no significant difference in SNR and CNR between the two algorithms (P>۰.۰۵).Conclusion: ASIR algorithm reduces the dose and image noise in head CT scan compared to the filtered back projection. In addition, using ASIR algorithm the image noise does not increase with lower mA.
کلیدواژه ها:
نویسندگان
Sahand Barzanjeh
School of Medicine, Hamadan University of Medical Sciences, Hamadan, Iran
Salman Jafari
Department of Radiology Technology, School of Paramedicine, Hamadan University of Medical Sciences, Hamadan, Iran
Kamaledin Hadei
Department of Radiology, School of Medicine, Hamadan University of Medical Sciences, Hamadan, Iran
Karim Ghazikhanlou sani
Department of Radiology Technology, School of Paramedicine, Hamadan University of Medical Sciences, Hamadan, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :