Image Quality and Entrance Surface Dose Evaluation of Lateral Cervical Spine: A Study Using Grid and Non-Grid Techniques
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 16، شماره: 2
سال انتشار: 1398
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 107
فایل این مقاله در 5 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-16-2_008
تاریخ نمایه سازی: 5 شهریور 1402
چکیده مقاله:
Introduction: The purpose of this study is to investigate the effects of grid and non-grid techniques in the lateral cervical spine radiography on image quality and entrance surface dose (ESD). Although image quality and radiation doses have been studied by researchers, there is still a dearth of information on image quality and patient dose with different techniques. Material and Methods: The radiographs of the lateral cervical spine were acquired by positioning the RANDO phantom abutting the erect bucky while using the grid and non-grid techniques. This study benefited from using a ۲۴ cm x ۳۰ cm Fuji standard cassette type imaging plate. A Leeds TOR test tool was utilized for relative comparison of image quality. The ESD of each examination was determined by using the optically stimulated luminescence dosimeter. Results: The increased kilovoltage (kVp) resulted in the reduction of ESD whether moving grid, stationary grid, or non-grid techniques were utilized. Significant differences in terms of contrast sensitivity and spatial resolution were indicated when comparing the grid technique to that of the non-grid technique (i.e., χ۲=۸ and ۵, ۱۶ respectively, p<۰.۰۵"> ). The results also indicated significant differences in ESD when using the moving grid, stationary grid, and non-grid techniques (i.e., χ۲=۷.۲, ۱۶p<۰.۰۵"> ). Conclusion: Significant differences in image quality and ESD were indicated when grid and non-grid techniques were used in the lateral cervical spine radiography. A non-grid with the highest appropriate kVp is recommended as the air gap acts as a grid, resulting in acceptable image quality with reduction in ESD.
کلیدواژه ها:
نویسندگان
Soo-Foon Moey
International Islamic University Malaysia
Nurfatin Nabilah Mohd Ramlee
International Islamic University Malaysia
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :