FDFD based measurement of permittivity using an open-ended coaxial probe

سال انتشار: 1402
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 139

فایل این مقاله در 13 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_JCESH-10-1_004

تاریخ نمایه سازی: 15 اسفند 1401

چکیده مقاله:

In this paper, ۲D-FDFD method is applied for measuring electrical properties of dielectric materials using an open-ended coaxial probe. A SMA connector with flange is used as the coaxial probe and the reflection coefficient from probe aperture in contact with dielectric is measured by a vector network analyzer. To convert the aperture-plane reflection coefficient to dielectric permittivity, first, the coaxial probe is modeled by the ۲D-FDFD method and then the genetic algorithm is employed to solve the inverse problem. The accuracy of the proposed method is investigated by measuring the dielectric properties of three known materials. The mean absolute percentage errors are below ۱۰% and the maximum absolute percentage error is below ۱۲%. Thus, good agreement between measured and actual values are observed.

نویسندگان

Toktam Shaterzadeh_Bojd

Computer and Communications Research Center, Ferdowsi University of Mashhad, Mashhad, Iran

Seyed Mohammad Saeed Majedi

Department of Electrical Engineering, Faculty of Engineering, Ferdowsi University of Mashhad, Mashhad, Iran