Pre-treatment Verification Performed with Electronic Portal Imager Device (EPID) and IMatriXX for ۲۰۴ Cancer Patients Treated with Intensity Modulated Radiotherapy (IMRT) – Phantom Based Study
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 18، شماره: 6
سال انتشار: 1400
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 224
فایل این مقاله در 8 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
این مقاله در بخشهای موضوعی زیر دسته بندی شده است:
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-18-6_010
تاریخ نمایه سازی: 20 آذر 1400
چکیده مقاله:
Introduction: State-of-art radiotherapy technique as Intensity Modulated Radiotherapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT) are being used to treat cancer with high accuracy. Verification of planned and delivered dose distribution is critical; in this study we evaluated quality assurance (QA) results and effectiveness of Electronic Portal Imaging Device (EPID) and IMatriXX. Material and Methods: Performance of EPID and IMatriXX was assessed with dose measurements using ionization chamber. Calibrated IMatriXX and EPID are used for pre-treatment patient-specific quality assurance (PSQA) for ۲۰۴ patients plans with IMRT treatment technique on LINAC. Dose image were compared for gamma evaluation (۳%/۳mm) and combination of three scalar parameters were assessed against EPID to quantify gamma results within region of interest; namely average g(gavg), maximum g(gmax) and Area Gamma<۱. Results: The g correlation comparisons yielded an average correlation of ۰.۹۹۱ for IMatriXX and ۰.۹۷۸ for EPID. The maximum gamma value is ۰.۹۹, while the minimum gamma is ۰.۸۷۲ for IMatriXX and ۰.۹۲۶ for EPID, which can be used as baseline. Our result suggests that EPID dosimetry, provides lower gamma correlation values than IMatriXX. Students Unpaired t-Test analysis was applied to two data sets. The calculated p-value ۰.۰۰۱ shows good correlation. Conclusion: The EPID and IMatriXX have significantly improved dosimetric properties, providing more sensitive, accurate pre-treatment PSQA. The result shows EPID can replace other ۲D dosimetry methods and ionization chamber measurements. It’s an efficient, sensitive and accurate dosimetry tool and is primary protocol of pre-treatment quality assurance.
کلیدواژه ها:
نویسندگان
Mahendra More
Department of Radiotherapy and Oncology, Rural Medical College, Pravara Institute of Medical Sciences (PIMS), Loni, Maharashtra; India
Vandana Jain
Department of Radiotherapy and Oncology, Rural Medical College, Pravara Institute of Medical Sciences (PIMS), Loni, Maharashtra; India
Om Prakash Gurjar
Government Cancer Hospital, Mahatma Gandhi Memorial Medical College, Indore, Madhya Pradesh; India
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :