A 790 nWLow-Noise Amplifier for Low Power Biopotentioal Signal Recording Systems

سال انتشار: 1398
نوع سند: مقاله کنفرانسی
زبان: انگلیسی
مشاهده: 646

فایل این مقاله در 5 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

ICMCONF01_021

تاریخ نمایه سازی: 19 اسفند 1398

چکیده مقاله:

In this paper an ultra-low-power amplifier for biomedical signal recording systems, especially for portable applications such as portable electrocardiogram (ECG) recording systems is presented. In this amplifier current reuse topology is used to achieve a lower input-referred voltage noise power density compared to a normal single-ended amplifier. Supply voltages are ±0.6 V, and the bias currents are 260 nA and 400 nA in the first and the second stages, respectively for reducing the power consumption. the proposed amplifier is simulated using Hspice and 0.18 μm TSMC CMOS technology. It is shown that the power consumotion of the proposed amplifieris as low as 790 nW. This amplifier has an open loop gain of about 100 dB and the total input-referred noise of 5.45μVrms in 7.1 KHz bandwidth and 3.7 μVrms in 200 Hz bandwidth for ECG recording.

نویسندگان

Fatemeh Karami Horestani

Electrical Engineering faculty shahid Beheshti University

Mohammad Eshghi

Electrical Engineering faculty shahid Beheshti University

Mohammadreza Yazdchi

Biomedical Engineering Department Faculty of Engineering, University of Isfahan,Iran

Mahmoud Norouzi

University of kashan Kashan,Isfahan,Iran