Calcium measurement using ion selective electrode based on a macromolecule ionophore

سال انتشار: 1396
نوع سند: مقاله کنفرانسی
زبان: انگلیسی
مشاهده: 438

نسخه کامل این مقاله ارائه نشده است و در دسترس نمی باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

ISPTC20_261

تاریخ نمایه سازی: 2 تیر 1397

چکیده مقاله:

A new PVC based membrane using benzo-18-crown-6 as an ionophore has been developed as a calcium ion selective electrode. The electrode comprises of ionophore (4%), PVC (30%) and plasticizer (63%). The electrode shows a linear dynamic response in the concentration range of 1×10−5 M to 1×10−1M with a Nernstian slope of 29.8 mV/decade and a detection limit as 4×10−6 M. The proposed sensor shows reasonably good selectivity with respect to alkali, alkaline earth and transition metal ions. It can be used in the pH range 5–12. The proposed electrode also was applied for the direct determination of Ca2+ cation in doubly distilled deionized water samples spiked with different concentrations calcium. This electrode was also used for determination of Ca2+ in real samples with a very good recovery.

نویسندگان

Sh Lovineh

Department of Chemistry, khouzestan Science and Research Branch, Islamic Azad University, Ahvaz, Iran- Department of chemistry, Ahvaz Branch, Islamic Azad University, Ahvaz, Iran

A Ghaemi

Department of chemistry, Ahvaz Branch, Islamic Azad University, Ahvaz, Iran