Determination of glucose by an electrochemical sensor based on molecularly imprinted polymer

  • سال انتشار: 1394
  • محل انتشار: اولین همایش ملی تکنیک های نوین در تجهیزات و مواد آزمایشگاهی صنعت نفت ایران
  • کد COI اختصاصی: IRANLABCO01_096
  • زبان مقاله: انگلیسی
  • تعداد مشاهده: 616
دانلود فایل این مقاله

نویسندگان

Mojdeh Alibakhshi

Electro analytical Chemistry Research Center, University of Science & Technology,Tehran, Iran

Mohammadreza Milani

Scientific Educational Research Center of Legal Medicine Organization, Tehran,Iran

چکیده

In this study a novel electrochemical sensor based on molecularly imprinted polymer has developed. The sensitive layer was prepared by electropolymerization of2mercaptobenzimidazole on a gold electrode in presence of glucose as template.The modifide electrode was washed with double distilled water for removal of template.The surface properties of prepared electrode were investigated by cyclic voltammetry technique. The detection limit was estimated 0.5mM corresponding to the 3standarddeviation of back ground current. The prepared sensor had selective response toward glucose concentration from 124 mM. The effect of interferences was also investigated.