Analysis of Ammonia in Human Breath Using Ion Mobility Spectrometry

  • سال انتشار: 1391
  • محل انتشار: دومین همایش ملی نفت ،گاز و پتروشیمی
  • کد COI اختصاصی: NCOGP02_198
  • زبان مقاله: انگلیسی
  • تعداد مشاهده: 1570
دانلود فایل این مقاله

نویسندگان

Elham Jazan

Department of Chemistry, Islamic Azad University, Shahreza Branch, ۳۱۱- ۸۶۱۴۵, Shahreza, Isfahan, Iran

Hadi Mirzaei

چکیده

In this study, ammonia in human breath is directly determined using ion mobility spectrometry (IMS). In this regard, the calibration curve was obtained after preparation of standard gaseous solutions of ammonia. The analytical results show the linear dynamic range between 10-810 ppb and the detection limit of 6.6 ppb. The relative standard deviation (RSD) was obtained 5.2, 3.4, and 3.0 for 290, 348, and 522 ppb, respectively. Then, amount of ammonia in breath of 8 healthy volunteers is measured and the values were between 236.0-1218 ppb. In addition, the amounts of ammonia were determined 1494-1553 ppb, for 2 kidney patients. In this work, it was shown that ion mobility spectrometry with corona discharge ionization source would be a successful technique for direct analysis of ammonia in human breath.

کلیدواژه ها

Ion Mobility Spectrometry, Ammonia, human breath