Investigation of Collapsed-cone Algorithm Accuracy in Small Fields and Heterogeneous Environments

  • سال انتشار: 1400
  • محل انتشار: مجله فیزیک و مهندسی پزشکی، دوره: 11، شماره: 2
  • کد COI اختصاصی: JR_JBPE-11-2_004
  • زبان مقاله: انگلیسی
  • تعداد مشاهده: 207
دانلود فایل این مقاله

نویسندگان

Razieh Zaghian

MSc, Medical Physics Department, Students Research Committee, Kermanshah University of Medical Sciences, Kermanshah, Iran

Abdolazim Sedighi Pashaki

MD, Radiation oncologist, Faculty of Medicine, Hamedan University of Medical Sciences, Hamedan, Iran

Abbas Haghparast

PhD, Department of Medical Physics, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran

Mohammad Hadi Gholami

PhD, Medical Physicist of Radiotherapy, Mahdieh Radiotherapy and Brachytherapy Charity Center, Hamedan, Iran

Mahdi Mohammadi

MSc, Medical Physics Department, Students Research Committee, Kermanshah University of Medical Sciences, Kermanshah, Iran

چکیده

Background: The use of small fields has increased by the emergence of advanced radiotherapy. Dose calculations of these fields are complex and challenging for many reasons such as lack of electrical equilibrium even in homogeneous environments, and this complexity will increase in presence of heterogeneity. According to the importance of delivery the accurate prescription dose to the target volume in the patient’s body, the dose calculation accuracy of used commercial algorithms in clinical treatment planning systems (TPS) should be evaluated. Objective: The present study aims to evaluate the accuracy of Collapsed-Cone dose measurement algorithm in Isogray treatment planning system.Material and Methods: In this analytical study, the measurements were made in tissue equivalent solid water phantom with lung and bone heterogeneities by Pinpoint dosimeter (۰.۰۱۵ cm۳ sensitive volume) in several radiation fields (۱×۱ to ۵×۵ cm۲). The phantoms were irradiated with ۶, ۱۰ and ۱۸ MV photon beams and finally, the results of experimental calculations were compared with treatment planning outputs. Results: In all setups, the maximum deviation occurred in the field of ۱×۱ cm۲. Then, the maximum deviation was observed for ۲×۲ cm۲ field size; however, it was up to ۵% for homogeneous water phantom and lung heterogeneity. In ۳×۳ cm۲ and larger fields, there was a good agreement between the results of the TPS and experimental dosimetry. The maximum deviation was observed in water-bone heterogeneity. Conclusion: This algorithm was able to pass the standard audit criteria, but it is better to be used more cautiously in bone heterogeneity, especially in low energies.

کلیدواژه ها

Radiotherapy, Dosimetry, Small Field, Algorithms, Phantoms, Imaging, Absorbed Dose

اطلاعات بیشتر در مورد COI

COI مخفف عبارت CIVILICA Object Identifier به معنی شناسه سیویلیکا برای اسناد است. COI کدی است که مطابق محل انتشار، به مقالات کنفرانسها و ژورنالهای داخل کشور به هنگام نمایه سازی بر روی پایگاه استنادی سیویلیکا اختصاص می یابد.

کد COI به مفهوم کد ملی اسناد نمایه شده در سیویلیکا است و کدی یکتا و ثابت است و به همین دلیل همواره قابلیت استناد و پیگیری دارد.