Verification of IMRT and RapidArc Localized Prostate Cancer Treatment Plans Using EPID and Delta۴ in vivo Dosimetry Methods
- سال انتشار: 1400
- محل انتشار: مجله سرطان خاورمیانه، دوره: 12، شماره: 4
- کد COI اختصاصی: JR_MISJ-12-4_012
- زبان مقاله: انگلیسی
- تعداد مشاهده: 47
نویسندگان
Physics Department, Faculty of Science, Mansoura University, Mansoura, Egypt
Physics Department, Faculty of Science, Mansoura University, Mansoura, Egypt
Clinical Oncology Department, Faculty of Medicine, Al Azhar University, Mansoura, Egypt
Clinical Oncology and Nuclear Medicine Department, Faculty of Medicine, Mansoura University, Mansoura, Egypt
Physics Department, Faculty of Science, Mansoura University, Mansoura, Egypt
چکیده
Background: Quality assurance for intensity modulated radiation therapy (IMRT) depends on the type of dosimetry system and its evaluation procedure. We can check both the intensity and distribution of each field as the required pretreatment verification with dosimetry systems. Method: Treatment verification for different plans (IMRT and RapidArc) applied on localized prostate cancer patients was done with electronic portal imager device (EPID), Delta۴). The EPID used was Varian aS۱۰۰۰ mounted on Varian (TrueBeam) Linac with gamma criteria set to ΔD=۳% and Δd=۳mm. RapidArc plans were designed by arcs (۱۷۹.۰o CCW to ۱۸۱.۰o and ۱۸۱.۰o CW to ۱۷۹.۰o) under the same gamma criteria of (ΔD=۳%, DTA=۳mm and Δ-index ≤۱), while the threshold dose was ۲۰%. Results: Evaluation analysis is passed for IMRT prostate plans with the area gamma ۰.۸ (was equal to ۲.۱%) / area gamma > ۱.۲ (was equal to ۰.۳%) and the average dose difference was ۰.۴۲CU. Delta۴ dosimetry system was assessed with RapidArc plan; the agreement between the measured and planned doses was ±۱% and gamma analysis resulted in ۱۰۰% data points with the same agreement conditions. Conclusion: Portal dosimetry provided a good verification of the treatment unit ability to deliver doses according to plan. For an IMRT field comprised of several subfields, it could give rise to much more errors. RapidArc plans were verified using Delta۴ system, which generated excellent dosimetry results. Periodic calibration was recommended for Delta۴ dosimetry system; radiation damage affected sensitivity by > ۱% every ۱kGy.کلیدواژه ها
Radiotherapy, Intensity-modulated, RapidArc, Radiometry, EPID, Delta۴اطلاعات بیشتر در مورد COI