Monte Carlo Dose Calculation of ۹۰ Sr/ ۹۰ Y Source in Water Phantom
- سال انتشار: 1387
- محل انتشار: مجله فیزیک پزشکی ایران، دوره: 5، شماره: 1
- کد COI اختصاصی: JR_IJMP-5-1_007
- زبان مقاله: انگلیسی
- تعداد مشاهده: 91
نویسندگان
Assistant Professor, Physics Dept., School of Sciences, Sabzevar Tarbiat Moallem University, Sabzevar, Iran
چکیده
Introduction: ۹۰ Sr/ ۹۰ Y source has been used for the intravascular brachytherapy to prevent coronary restenosis in the patients who have undergone angioplasty. The aim of this research is to determine the dose distribution of ۹۰ Sr/ ۹۰ Y source in a water phantom. Materials and Methods: In the present work, MCNP code has been applied to calculate the dose distribution around a ۳ cm length of ۹۰ Sr/ ۹۰ Y source in a ۳۰×۳۰×۳۰ cm ۳ water phantom. Also, the exact geometry of the source has been used in this simulation. Tally *F۸:e which is suitable for beta ray dosimetry has been evaluated with less than %۵ relative error in a sphere having ۰.۲ mm radius. Results: The isodose curve for ۱۰, ۲۰, ۴۰, and ۹۰% depth dose (PDD) were derived based on the calculated dose curves along the parallel and perpendicular axis to the source. Discussion and Conclusion: The results obtained in this work are in a good agreement with the experimental result published by Buckley et al. and the International Atomic Energy Agency (IAEA) report in a water phantom. Therefore, the result of this research can be used in the intravascular brachytherapy.کلیدواژه ها
۹۰ Sr/ ۹۰ Y Source, Intravascular Brachytherapy, MCNP۴C Code, Water Phantom, Isodose Curvesاطلاعات بیشتر در مورد COI
COI مخفف عبارت CIVILICA Object Identifier به معنی شناسه سیویلیکا برای اسناد است. COI کدی است که مطابق محل انتشار، به مقالات کنفرانسها و ژورنالهای داخل کشور به هنگام نمایه سازی بر روی پایگاه استنادی سیویلیکا اختصاص می یابد.
کد COI به مفهوم کد ملی اسناد نمایه شده در سیویلیکا است و کدی یکتا و ثابت است و به همین دلیل همواره قابلیت استناد و پیگیری دارد.