Optimized design and implementation of low-cost, sensitive and versatile Vibrating Sample Magnetometer

  • سال انتشار: 1391
  • محل انتشار: بیستمین کنفرانس مهندسی برق ایران
  • کد COI اختصاصی: ICEE20_418
  • زبان مقاله: انگلیسی
  • تعداد مشاهده: 1373
دانلود فایل این مقاله

نویسندگان

Nahid Hosseini

Electrical Engineering Department, Sharif University of Technology, Tehran, Iran

Somayeh Khiabani

Farrokh Sarreshtedari

Mehdi Fardmanesh

چکیده

Geometry of different electromagnetic parts of Vibrating Sample Magnetometer (VSM) is a major factor of VSM output voltage. In this paper design, geometryoptimization and implementation of a sensitive, versatile and relatively inexpensive VSM system has been described. Considering all effective geometrical parameters in the systemoutput voltage, the induced voltage in the pick-up coils has been obtained analytically and the geometry of the pick-up coils hasbeen optimized for maximum output induced voltage. Two opposite series of pick-up coils, piezoelectric based vibrationsystem and required electronic circuits have been designed and the VSM system has been implemented. The implemented optimized VSM system has been used to reveal magnetization loop of ferromagnetic materials

کلیدواژه ها

Ferromagnetic, Magnetization loop, Pickup coil, VSM